How do I use the new stickers with my Delvotest Go?

1. Gather materials

2. Prepare the ampoules as per instructions in the app. See images - how to open the ampoule - pierce the ampoule using the corner of the grid.

3. Use the pipette to aspire milk and add the milk 0.1ml = 100 ul into the ampoule. Do not pierce (touch) the purple agar gel.

4. Take one sticker from the provided sheet and carefully place it on top to seal the ampoule.

5. Likewise prepare the extra ampoules. Number the ampoules to identify the samples.

Note : it is advised to write numbers on the stickers, but not on the ampoules.

6. When ready in the app add all ampoules into the slots. If reference ampoule is needed, add it into slot-1.

7. Close the lid.

8. Add comments in the app available for your other samples.

Next article
In the upcoming articles, we will explain what calibration files are generated and how you can access them . If you're following the Installation and Setup flow, now that you've set up your calibration cycles, it's time to start with your first calibration—position. In the next article, we will guide you through what you'll need and how to do this.
Before you start
Published: 10 November 2025
Important
Although diligent care has been used to ensure that the information provided herein is accurate, nothing contained herein can be construed to imply any representation or warranty for which we assume legal responsibility, including without limitation any warranties as to the accuracy, currency or completeness of this information or of noninfringement of third party intellectual property rights. The content of this document is subject to change without further notice. This document is non-controlled and will not be automatically replaced when changed. Please contact us for the latest version of this document or for further information. Since the user's product formulations, specific use applications and conditions of use are beyond our control, we make no warranty or representation regarding the results which may be obtained by the user. It shall be the responsibility of the user to determine the suitability of our products for the user's specific purposes and the legal status for the user's intended use of our products.©DSM Food Specialties B.V. | Trade Register Number 27235314